Authors: Athanasios Katsargyris Kyriakos Oikonomou Umberto M Bracale Eric L G Verhoeven
Publish Date: 2012/05/09
Volume: 36, Issue: 2, Pages: 536-539
Abstract
The new C3 Gore Excluder delivery system enables both up/downward and rotational repositioning of the device before complete deployment This contributes to more precise proximal landing and permits facilitation of the contralateral gate cannulation During separate deployment the position of the ipsilateral limb can also be readjusted We have used the modified C3 delivery system in more than 50 patients and in most cases we were able to utilize the repositioning options of the device to achieve optimal fixation and sealing However we present a case where our attempt to readjust the position of the ipsilateral limb led to upward migration of the main body with coverage of the left renal artery origin The latter was secured with a bare stent implantation